Objectives:
To evaluate the feasibility, safety, and efficacy of implantation of the new Nit Occlud ASD-R® (NOASD-R) device for percutaneous closure of ostium secundum atrial septal defects (ASD-OS).
Background:
Device catheter implantation has become the method of choice for most patients with ASD-OS. No single device has proven to be ideal for this type of procedure. The NOASD-R has a distinct design that may help to overcome limitations of other devices.
Methods:
A prospective, single arm, observational study including all consecutive patients receiving the NOASD-R device for ASD-OS closure between October 2011 and September 2013 was performed. Patient selection, device design, deployment technique, complications, and procedural outcomes were evaluated.
Results:
Seventy-four patients underwent attempted transcatheter ASD-OS closure using the NOASD-R device. Implantation of the occluder was successful in 73 patients (98.6%). The majority of patients were female (79.5%) with a median age of 17.2 years (range: 2-74). A 2-D transthoracic color-Doppler echocardiogram (TTE) obtained at the 3 or 6 month follow-up visit showed complete occlusion of the ASD-OS in 72/73 patients (98.6%). At a mean follow-up interval of 11.4 ± 6.8 months there have been no episodes of late device embolization, cardiac perforation or erosion, endocarditis, thromboembolism, wire fracture, embolic neurologic events, or death.
Conclusions:
We report the first worldwide clinical experience using the NOASD-R device for ASD-OS closure. The procedure was feasible, with a high rate of successful implantations, and safe. High ASD-OS closure rates and no complications were encountered during short-term follow-up. © 2014 Wiley Periodicals, Inc.