Related Experiment Video
Updated: Apr 29, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Applications of memory alloy stent in vertebral fractures
Yang Yimin1, Zhang Zhi1, Ren ZhiWei1
1Department of Orthopedics, First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China (mainland).
Background:
The aim of this study was to evaluate the feasibility of treating vertebral compression fractures using an autonomously developed nitinol memory alloy vertebral stent.
Material And Methods:
Thoracolumbar vertebral specimens from adult human cadavers were made into models of compression fractures. The models were divided into group A, which received percutaneous kyphoplasty (PKP), balloon dilation, and nitinol memory alloy vertebral stent implantation (PKP + nitinol stent group); group B, which received percutaneous vertebroplasty (PVP) and direct implantation of a nitinol memory alloy vertebral stent (PVP + nitinol stent group); and group C, which received PKP, balloon dilation, and bone cement vertebroplasty (PKP + polymethylmethacrylate (PMMA) group). Vertebral heights were measured before and after the surgery and the water bath incubation to compare the impact of the 3 different surgical approaches on reducing vertebral compression.
Results:
The 3 surgical groups could all significantly restore the heights of compressed vertebral bodies. The vertebral heights of the PKP + nitinol stent group, PVP + nitinol stent group, and PKP + PMMA group were changed from the preoperative levels of (1.59±0.08) cm, (1.68±0.08) cm, and (1.66±0.11) cm to the postoperative levels of (2.00±0.09) cm, (1.87±0.04) cm, and (1.99±0.09) cm, respectively. After the water bath, the vertebral heights of each group were changed to (2.10±0.07) cm, (1.98±0.09) cm, and (2.00±0.10) cm, respectively. Pairwise comparison of the differences between the preoperative and postoperative vertebral heights showed that group A and group B differed significantly (P=0.000); group B and group C differed significantly (P=0.003); and group A and group C had no significant difference (P=0.172). Pairwise comparison of the differences in the vertebral heights before and after the water bath showed that group A and group C differed significantly (P=0.000); group B and group C differed significantly (P=0.000); and group A and group B had no significant difference (P=0.157).
Conclusions:
The nitinol memory alloy stents can effectively support and reduce the compression of vertebral endplates and can be used to treat vertebral compression fractures without neurological symptoms.
More Related Videos
09:24A Novel Vertebral Stabilization Method for Producing Contusive Spinal Cord Injury
Published on: January 5, 2015
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015