Closed-Loop Control Performance of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System in a Feasibility Study
Daniel A Finan1, Thomas W McCann2, Linda Mackowiak2
1Animas Corporation, West Chester, PA, USA dfinan@its.jnj.com.
Background:
This feasibility study investigated the insulin-delivery characteristics of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System-an automated insulin delivery device-in participants with type 1 diabetes.
Methods:
Thirteen adults with type 1 diabetes were enrolled in this nonrandomized, uncontrolled, clinical-research-center-based feasibility study. The HHM System comprised a continuous subcutaneous insulin infusion pump, a continuous glucose monitor (CGM), and a model predictive control algorithm with a safety module, run on a laptop platform. Closed-loop control lasted approximately 20 hours, including an overnight period and two meals.
Results:
When attempting to minimize glucose excursions outside of a prespecified target zone, the predictive HHM System decreased insulin infusion rates below the participants' preset basal rates in advance of below-zone excursions (CGM < 90 mg/dl), and delivered 80.4% less insulin than basal during those excursions. Similarly, the HHM System increased infusion rates above basal during above-zone excursions (CGM > 140 mg/dl), delivering 39.9% more insulin than basal during those excursions. Based on YSI, participants spent a mean ± standard deviation (SD) of 0.2 ± 0.5% of the closed-loop control time at glucose levels < 70 mg/dl, including 0.3 ± 0.9% for the overnight period. The mean ± SD glucose based on YSI for all participants was 164.5 ± 23.5 mg/dl. There were nine instances of algorithm-recommended supplemental carbohydrate administrations, and there was no severe hypoglycemia or diabetic ketoacidosis.
Conclusions:
Results of this study indicate that the current HHM System is a feasible foundation for development of a closed-loop insulin delivery device.
More Related Videos
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
Related Concept Videos
Hypoglycemia and Glucagon
Hypoglycemia
Feedback Loops
Hyperosmolar Hyperglycemic State
Hyperglycemia
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...