Surface quality of human corneal lenticules after SMILE assessed using environmental scanning electron microscopy
Purpose:
To assess the cut quality of the anterior and posterior surfaces of intrastromal refractive lenticules removed during small incision lenticule extraction (SMILE).
Methods:
The VisuMax femtosecond laser (500 kHz; Carl Zeiss Meditec, Dublin, CA) was used to perform SMILE on 8 eyes of 5 individuals to correct only myopia (no cylinder). The cut energy index was 26 (equivalent to an energy of 130 nJ) with a 2.5 × 2.5 μm spot/track separation. The lenticule diameter was 6.5 mm with a minimum edge thickness of 15 μm and the cap diameter was 7.3 mm with an intended thickness of 120 μm. After laser treatment, the lenticule was loosened with a spatula and removed with forceps. The extracted lenticules were placed in 2% formalin and sent for imaging with an environmental scanning electron microscope. Images of the anterior and posterior surfaces of the lenticules were obtained at multiple magnifications (100×, 250×, and 500×). Surface quality was evaluated by an investigator who specializes in electron microscopy using three criteria: overall surface regularity, percent of surface irregularity, and position of irregular area.
Results:
Both the anterior and posterior surfaces of the extracted lenticules were smooth and absent of surface irregularities. The cut edges also appeared uniform. Jagged edges were seen in several images, but were clearly caused by the forceps during extraction.
Conclusions:
Using the VisuMax laser to perform SMILE produces smooth cuts absent of surface irregularities.
More Related Videos
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
08:38A Standardized Method for the Analysis of Liver Sinusoidal Endothelial Cells and Their Fenestrations by Scanning Electron Microscopy
Published on: April 30, 2015
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Confocal Fluorescence Microscopy
Preparation of Samples for Electron Microscopy
Overview of Electron Microscopy