Output calibration in solid water for high energy photon beams
1Department of Radiation Oncology, University of Chicago, Illinois 60637.
Medical Physics
|March 1, 1989
Summary
This study validates the AAPM Protocol
Area of Science:
- Medical Physics
- Radiation Dosimetry
Background:
- The AAPM Protocol recommends water, polystyrene, or acrylic for high-energy photon beam output measurements.
- Conversion of dose measurements from these media to water requires specific coefficients.
Purpose of the Study:
- To investigate the accuracy of calculating restricted mass stopping powers and mass energy absorption coefficients for a water-equivalent solid.
- To validate the AAPM Protocol's methodology for determining dose to water from various measurement media.
Main Methods:
- Output measurements were performed using a Farmer type chamber.
- Measurements were conducted in four materials for Co-60 and 4-24 MV photon beams, adhering to the AAPM Protocol.
- Calculated restricted mass stopping powers and mass energy absorption coefficients were utilized.
Main Results:
- Scaled dose to water measurements across different media agreed within 1%.
- The analysis confirmed the accuracy of the published coefficients for the water-equivalent solid.
- The findings support the AAPM Protocol's approach for dose-to-water determination.
Conclusions:
- The AAPM Protocol's methodology is accurate for determining dose to water from polystyrene, acrylic, and a water-equivalent solid.
- The validated coefficients enhance the precision of radiation dosimetry in high-energy photon beams.