Related Experiment Video
Updated: Apr 27, 2026

Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
Published on: January 30, 2017
Multicenter clinical evaluation of the novel Alere™ i Influenza A&B isothermal nucleic acid amplification test
Jeremiah Bell1, Aleta Bonner2, Daniel M Cohen3
1Children's Mercy Hospitals and Clinics, Kansas City, MO, United States; University of Missouri Kansas City School of Medicine, Kansas City, MO, United States.
Background:
Rapid detection of influenza infection is important for patient management and timely anti-viral therapy. Rapid antigen detection tests for influenza have inferior sensitivity when compared to nucleic acid-based amplification tests. An isothermal nucleic acid amplification test that offers the potential for rapid molecular testing at the clinical point-of-care with simple equipment can improve influenza detection rates.
Objectives:
To evaluate the performance of Alere™ i Influenza A&B isothermal nucleic acid amplification test to detect influenza A and B in comparison to viral cell culture as reference method.
Study Design:
A prospective, multicenter, clinical study to evaluate the clinical performance of the Alere™ i Influenza A&B assay in a point-of-care setting using prospectively enrolled specimens from both children and adults was conducted in seven sites.
Results:
In comparison with viral cell culture, the overall sensitivity and specificity of the Alere™ i Influenza A&B assay was 97.8% and 85.6% for the detection of influenza A, and 91.8% and 96.3% for the detection of influenza B, respectively. Following resolution of discrepant results by real-time RT-PCR the sensitivity and specificity of the Alere™ i Influenza A&B assay improved to 99.3% and 98.1% for influenza A, and 97.6% and 100% for influenza B, respectively.
Conclusions:
The Alere™ i Influenza A&B isothermal nucleic acid amplification test is an ideal point-of-care test for influenza detection in children and adults due to its high sensitivity and specificity and ability to generate results within 15 min from specimen receipt.