Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of
Sohi Bae1, Myung-Joon Kim, Choon-Sik Yoon
1Department of Radiology and Research Institute of Radiological Science, Severance Children's Hospital, Yonsei University College of Medicine, 50-1, Yonsei-ro, Seodaemun-gu, 120-752, Seoul, South Korea.
Background:
Since children are more radio-sensitive than adults, there is a need to minimize radiation exposure during CT exams.
Objective:
To evaluate the effects of adaptive statistical iterative reconstruction (ASIR) on radiation dose reduction, image quality and diagnostic accuracy in pediatric abdominal CT.
Materials And Methods:
We retrospectively reviewed the abdominal CT examinations of 41 children (24 boys and 17 girls; mean age: 10 years) with a low-dose radiation protocol and reconstructed with ASIR (the ASIR group). We also reviewed routine-dose abdominal CT examinations of 41 age- and sex-matched controls reconstructed with filtered-back projection (control group). Image quality was assessed objectively as noise measured in the liver, spleen and aorta, as well as subjectively by three pediatric radiologists for diagnostic acceptability using a four-point scale. Radiation dose and objective image qualities of each group were compared with the paired t-test. Diagnostic accuracy was evaluated by reviewing follow-up imaging studies and medical records in 2012 and 2013.
Results:
There was 46.3% dose reduction of size-specific dose estimates in ASIR group (from 13.4 to 7.2 mGy) compared with the control group. Objective noise was higher in the liver, spleen and aorta of the ASIR group (P < 0.001). However, the subjective image quality was average or superior in 84-100% of studies. Only one image was subjectively rated as unacceptable by one reviewer. There was only one case with interpretational error in the control group and none in the ASIR group.
Conclusion:
Use of the ASIR technique resulted in greater than a 45% reduction in radiation dose without impairing subjective image quality or diagnostic accuracy in pediatric abdominal CT, despite increased objective image noise.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...