Related Experiment Video
Updated: Apr 26, 2026

A Simple Fractionated Extraction Method for the Comprehensive Analysis of Metabolites, Lipids, and Proteins from a Single Sample
Published on: June 1, 2017
An automatic versatile system integrating solid-phase extraction with ultra-high performance liquid
Qisheng Zhong1, Xiongxiong Qiu1, Caiyong Lin1
1Shimadzu Global COE for Application & Technical Development, Guangzhou, 510010, China.
Abstract:
An automatic versatile system which integrated solid phase extraction (SPE) with ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) was developed. Diverse commercial SPE columns can be used under an ambient pressure in this online system realized by a dual-dilution strategy. The first dilution enabled the direct injection of complex samples with minimal pretreatment, and the second dilution realized direct introduction of large volume of strong eluent into the UHPLC column without causing peak broadening or distortion. In addition, a post-column compensation mode was also designed for the matrix-effects evaluation. The features of the online system were systematically investigated, including the dilution effect, the capture of desorption solution, the column-head stacking effect and the system recovery. Compared with the offline UHPLC system, this online system showed significant advantages such as larger injection volume, higher sensitivity, shorter analysis time and better repeatability. The feasibility of the system was demonstrated by the direct analysis of three auxins from different plant tissues, including leaves of Dracaena sanderiana, buds and petals of Bauhinia. Under the optimized conditions, the whole analysis procedure took only 7min. All the correlation coefficients were greater than 0.9987, the limits of detection and the limits of quantitation were in the range of 0.560-0.800ng/g and 1.80-2.60ng/g, respectively. The recoveries of the real samples ranged from 61.0 to 117%. Finally, the post-column compensation mode was applied and no matrix-effects were observed under the analysis conditions. The automatic versatile system was rapid, sensitive and reliable. We expect this system could be extended to other target analytes in complex samples utilizing diverse SPE columns.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
09:36Author Spotlight: Discovering New Alkaloids in Plants with Advanced Mass Spectrometry Techniques
Published on: March 8, 2024
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Instrumentation