An Improvised "Blow Glove" Device Produces Similar PEP Values to a Commercial PEP Device: An Experimental Study
Yaakov Dagan1, Itay Wiser2, Oren Weissman3
1Physiotherapy Service.
Background:
Postoperative positive expiratory pressure (PEP) therapy promotes increased lung volume, secretion clearance, and improved oxygenation. Several commercial devices exist that produce recommended PEP values (10-20 cmH2O) when the patient breathes through a fixed orifice resistor. It was hypothesized that an inexpensive, improvised "blow glove" device would produce similar PEP values over a wider range of expiration volumes and flow rates.
Methods:
PEP for different expiration volumes (400-2000 mL) and expiratory flow rates (10-80 L/min) was compared between a commercial PEP device (Resistex, Mercury Medical, Clearwater, FL) and an improvised "blow glove" device, recorded by a Vela ventilator (CareFusion, San Diego, CA). Dynamics in positive end expiratory pressure (PEEP) values were evaluated following five consecutive expirations. The "blow glove" device was evaluated using various glove compositions and sizes.
Results:
The improvised "blow glove" device produced a significantly higher rate of PEP values in the recommended range than the Resistex device (88.9% vs. 20%, p<0.0001). No significant difference was observed between small and large glove sizes (88.9% vs. 82.9%, p>0.05), but the powdered latex glove showed a significantly higher rate of PEP values in the recommended range than the powder-free latex glove (88.9% vs. 44.4%, p<0.001).
Conclusions:
A "blow glove" PEP device using a powdered latex glove produces PEP values in the recommended range over a wider spectrum of expiratory flow rates and expiration volumes than a commercial PEP device.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...