Related Experiment Video
Updated: Apr 25, 2026

Endotracheal Intubation of Rabbits Using a Polypropylene Guide Catheter
Published on: November 13, 2017
Fluoroscopic guidance for placing a double lumen endotracheal tube in adults
Emile Calenda1, Jean Marc Baste2, Ridha Hajjej1
1Department of Anesthesia, Rouen University Hospital, Centre Hospitalier Universitaire de Rouen, Rouen, France.
Objective:
The aim of this study was to assess the right placement of the double lumen endotracheal tube with fluoroscopic guidance, which is used in first intention prior to the fiberscope in our institution.
Methods:
This was a prospective observational study. The study was conducted in vascular and thoracic operating rooms. We enrolled 205 patients scheduled for thoracic surgery, with ASA physical statuses of I (n = 37), II (n = 84), III (n = 80), and IV (n = 4). Thoracic procedures were biopsy (n = 20), wedge (n = 34), culminectomy (n = 6), lobectomy (n = 82), pneumonectomy (n = 4), sympathectomy (n = 9), symphysis (n = 47), and thymectomy (n = 3). The intubation with a double lumen tube was performed with the help of a laryngoscope. Tracheal and bronchial balloons were inflated and auscultation was performed after right and left exclusions. One shot was performed to locate the position of the bronchial tube and the hook. Fluoroscopic guidance was used to relocate the tube in case of a wrong position. When the fluoroscopic guidance failed to position the tube, a fiberscope was used. Perioperative collapse of the lung was assessed by the surgeon during the surgery.
Results:
Correct fluoroscopic image was obtained after the first attempt in 58.5% of patients therefore a misplaced position was encountered in 41.5%. The fluoroscopic guidance allowed an exact repositioning in 99.5% of cases, and the mean duration of the procedure was 8 minutes. A fiberscope was required to move the hook for one patient. We did not notice a moving of the double lumen endotracheal tube during the surgery. The surgeon satisfaction was 100%.
Conclusion:
The fluoroscopy evidenced the right position of the double lumen tube and allowed a right repositioning in 99.5% of patients with a very simple implementation.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...