Quantitative digital subtraction arteriography with a calibration catheter
K H Barth1, P W Kremers, D Lindisch
1Department of Radiology, Georgetown University Medical Center, Washington, DC 20007.
Cardiovascular and Interventional Radiology
|September 1, 1989
Summary
Digital subtraction arteriography (DSA) measurements of abdominal arteries showed acceptable accuracy for clinical use. A calibration catheter with metallic markers improved measurement reliability in this quantitative imaging study.
Area of Science:
- Vascular Imaging
- Medical Device Technology
- Radiology
Background:
- Accurate measurement of arterial diameters is crucial for diagnosing and managing vascular diseases.
- Digital subtraction arteriography (DSA) is a common imaging modality, but its quantitative accuracy requires validation.
Purpose of the Study:
- To evaluate the accuracy of quantitative measurements of abdominal arteries using digital subtraction arteriography (DSA).
- To assess the impact of a calibration catheter with metallic markers on measurement accuracy.
Main Methods:
- Quantitative intraarterial digital subtraction arteriography (DSA) was performed using a calibration catheter with three distal metallic ring markers.
- Measurements of 54 vessel diameters (abdominal aorta, renal, lumbar, iliac arteries) were compared against direct film arteriograms.
- Diameter measurements were performed by both human observers and computer software.
Main Results:
- Measurement deviations on DSA images ranged from 8-13% for vessels <5 mm and 2-6% for vessels >=5 mm, compared to film standards.
- Projectional artifacts contributed 3% or less error to measurements.
- The calibration catheter served as a reliable internal marker.
Conclusions:
- Quantitative DSA provides acceptable accuracy for measuring abdominal artery diameters, with variations dependent on vessel size.
- Understanding these measurement variations is essential for establishing clinical error tolerances.
- The use of a calibration catheter enhances the reliability of DSA measurements.