Do the flags related to immature granulocytes reported by the Sysmex XE-5000 warrant a microscopic slide review?

Heidi Eilertsen1, Tor-Arne Hagve2

  • 1From the Division of Diagnostics and Technology, Akershus University Hospital, Lørenskog, Norway; Faculty of Health Sciences, Oslo, and Akershus University College of Applied Sciences, Oslo, Norway; and eilertsen@hioa.no.

Abstract

Insights

The Sysmex XE-5000 "Imm Gran?" flag shows poor reliability for immature granulocytes (IGs). Automated IG counts are reproducible but higher than manual counts, suggesting reporting without initial slide review.

Area of Science:

  • Hematology
  • Clinical Chemistry
  • Laboratory Medicine

Background:

  • Automated hematology analyzers like the Sysmex XE-5000 provide immature granulocyte (IG) counts.
  • The "Imm Gran?" flag is intended to signal potentially unreliable IG results.

Purpose of the Study:

  • To evaluate the clinical utility of the "Imm Gran?" flag on Sysmex XE-5000 instruments.
  • To assess the analytical and diagnostic performance of automated immature granulocyte (IG) measurements.

Main Methods:

  • Analyzed 408 samples across three Sysmex XE-5000 instruments.
  • Assessed reproducibility of IG counts and "Imm Gran?" flag reports.
  • Compared automated IG counts and flag data against manual differential counts.

Main Results:

  • "Imm Gran?" flag reproducibility between instruments was poor (κ = 0.75–0.80).
  • Bands were the primary cause of the "Imm Gran?" flag; its role in blast detection was minimal.
  • Interinstrument reproducibility for IG counts was high (ICC = 0.99).
  • Automated IG counts were consistently higher (36%–55%) than manual counts, with increasing variability at higher IG levels.

Conclusions:

  • The "Imm Gran?" flag exhibits poor analytical quality and offers limited value in identifying blasts.
  • Automated IG counts demonstrate high reproducibility.
  • Recommended reporting automated IG counts directly, bypassing initial microscopic review.