Inter- and intravein differences in cardiac output with cardiac resynchronization pacing using a multipolar LV pacing
Michael A Jones1, Raj Khiani, Paul Foley
1From the Cardiology Department, John Radcliffe Hospital, Oxford University Hospitals NHS Trust, Headington, Oxfordshire, UK.
Background:
Quadripolar left ventricular pacing leads permit a variety of pacing configurations from different sites within a coronary vein. There may be advantages to selecting a specific pacing vector. This study examines whether the range of cardiac outputs obtained at cardiac resynchronization therapy (CRT) implantation is greater between different poles within a vein, or greater between two different veins.
Methods And Results:
The cardiac index (CI, L/min/m(2) ) was measured during CRT implantation using a noninvasive cardiac output monitor (NICOM™, Cheetah Medical Inc., Newton Center, MA, USA) and a quadripolar left ventricle (LV) lead, in 22 patients with sinus rhythm. CI was recorded during right atrial-biventricular pacing at 70/min with fixed atrioventricular and ventriculo-ventricular delay, from each LV electrode in one vein, and then from an alternate vein. Phrenic nerve stimulation (PNS) occurred in nine of 15 posterior and three of 21 anterior veins (P = 0.005). At least one electrode in each vein had no PNS. The mean (standard deviation [SD]) difference between best and worst CI within any one vein was 13.1% (±9%). The mean (SD) difference between the best CI in one vein compared to the other was 9.8% (±8%; P = 0.043). In 16 of 22 patients, the range of CI was greater between poles within one vein, rather than between two veins (best of one vein compared to best from the other). In four of 22 patients, the range was greater between veins (P = 0.0003).
Conclusion:
A greater range of CI is found within a single vein than between two different veins. This finding has implications both for the approach to implant technique and postimplant programming and optimization.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization