Related Experiment Video
Updated: Apr 20, 2026

Clinical Imaging of Microwave Mammography
Published on: November 14, 2025
A method for safety testing of radiofrequency/microwave-emitting devices using MRI
Leeor Alon1,2,3,4, Gene Y Cho1,2,4, Xing Yang
1The Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University, School of Medicine, New York, New York, USA.
Purpose:
Strict regulations are imposed on the amount of radiofrequency (RF) energy that devices can emit to prevent excessive deposition of RF energy into the body. In this study, we investigated the application of MR temperature mapping and 10-g average specific absorption rate (SAR) computation for safety evaluation of RF-emitting devices.
Methods:
Quantification of the RF power deposition was shown for an MRI-compatible dipole antenna and a non-MRI-compatible mobile phone via phantom temperature change measurements. Validation of the MR temperature mapping method was demonstrated by comparison with physical temperature measurements and electromagnetic field simulations. MR temperature measurements alongside physical property measurements were used to reconstruct 10-g average SAR.
Results:
The maximum temperature change for a dipole antenna and the maximum 10-g average SAR were 1.83°C and 12.4 W/kg, respectively, for simulations and 1.73°C and 11.9 W/kg, respectively, for experiments. The difference between MR and probe thermometry was <0.15°C. The maximum temperature change and the maximum 10-g average SAR for a cell phone radiating at maximum output for 15 min was 1.7°C and 0.54 W/kg, respectively.
Conclusion:
Information acquired using MR temperature mapping and thermal property measurements can assess RF/microwave safety with high resolution and fidelity.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging