Flow simulations and validation for the first cohort of patients undergoing the Y-graft Fontan procedure
Weiguang Yang1, Frandics P Chan2, V Mohan Reddy3
1Department of Pediatrics (Cardiology), Stanford University, Palo Alto, Calif.
Objectives:
In this study, with the use of computational fluid dynamics, we evaluate the postoperative hemodynamic performance of the first cohort of patients undergoing a handcrafted Y-graft Fontan procedure and validate simulation predictions of hepatic blood flow distribution against in vivo clinical data.
Methods:
An 18-12 × 2-mm handcrafted Y-graft modification of the Fontan procedure was performed in 6 patients. Early (at the time of discharge) and 6-month postoperative 3-dimensional magnetic resonance imaging data were collected. Patient-specific models were constructed for flow simulations.
Results:
Hepatic blood flow distribution varied among patients. Lung perfusion data (n = 3) showed good agreement with simulations. Postoperative asymmetry in hepatic blood flow distribution was reduced 6 months postoperatively. In 1 patient, low wall shear stress was found in the left limb of the Y-graft, corresponding to the location of subsequent thrombosis in the patient.
Conclusions:
The credibility and accuracy of simulation-based predictions of postoperative hepatic flow distribution for the Fontan surgery have been validated by in vivo lung perfusion data. The performance of the Y-graft design is highly patient-specific. The anastomosis location is likely the most important factor influencing hepatic blood flow distribution. Although the development of thrombosis is multifactorial, the occurrence in 1 patient suggests that simulations should not solely consider the hepatic blood flow distribution but also aim to avoid low wall shear stress in the limbs.