Simultaneous Affinity Enrichment of Two Post-Translational Modifications for Quantification and Site Localization
Published on: February 27, 2020
Expanding the linear dynamic range for quantitative liquid chromatography-high resolution mass spectrometry utilizing
Hanghui Liu1, Lily Lam1, Lin Yan1
1Senomyx Inc. 4767 Nexus Centre Dr., San Diego, CA 92121, USA.
Abstract:
The linear dynamic range (LDR) for quantitative liquid chromatography-mass spectrometry can be extended until ionization saturation is reached by using a number of target isotopologue ions in addition to the normally used target ion that provides the highest sensitivity. Less abundant isotopologue ions extend the LDR: the lower ion abundance decreases the probability of ion detector saturation. Effectively the sensitivity decreases and the upper limit of the LDR increases. We show in this paper that the technique is particularly powerful with a high resolution time of flight mass spectrometer because the data for all ions are automatically acquired, and we demonstrated this for four small organic molecules; the upper limits of LDRs increased by 25-50 times.