Sudkate Chaiyo1, Orawon Chailapakul2, Weena Siangproh1
1Department of Chemistry, Faculty of Science, Srinakharinwirot University, Thailand.
Abstract:
A highly sensitive determination of mercury in the presence of Cu(II) using a boron-doped diamond (BDD) thin film electrode coupled with sequential injection-anodic stripping voltammetry (SI-ASV) was proposed. The Cu(II) was simultaneously deposited with Hg(II) in a 0.5 M HCl supporting electrolyte by electrodeposition. In presence of an excess of Cu(II), the sensitivity for the determination of Hg(II) was remarkably enhanced. Cu(II) and Hg(II) were on-line deposited onto the BDD electrode surface at -1.0 V (vs. Ag/AgCl, 3 M KCl) for 150 s with a flow rate of 14 μL s(-1). An anodic stripping voltammogram was recorded from -0.4 V to 0.25 V using a frequency of 60 Hz, an amplitude of 50 mV, and a step potential of 10 mV at a stopped flow. Under the optimal conditions, well-defined peaks of Cu(II) and Hg(II) were found at -0.25 V and +0.05 V (vs. Ag/AgCl, 3 M KCl), respectively. The detection of Hg(II) showed two linear dynamic ranges (0.1-30.0 ng mL(-1) and 5.0-60.0 ng mL(-1)). The limit of detection (S/N=3) obtained from the experiment was found to be 0.04 ng mL(-1). The precision values for 10 replicate determinations were 1.1, 2.1 and 2.9% RSD for 0.5, 10 and 20 ng mL(-1), respectively. The proposed method has been successfully applied for the determination of Hg(II) in seawater, salmon, squid, cockle and seaweed samples. A comparison between the proposed method and an inductively coupled plasma optical emission spectrometry (ICP-OES) standard method was performed on the samples, and the concentrations obtained via both methods were in agreement with the certified values of Hg(II), according to the paired t-test at a 95% confidence level.
More Related Videos
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Electrodeposition
Electrodeposition can...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetric Techniques: Cyclic Voltammetry
Electrogravimetric Analysis: Overview
To test the completeness of the...