Related Experiment Video
Updated: Apr 20, 2026

Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
Pulse oximetry measures a lower heart rate at birth compared with electrocardiography
Jeroen J van Vonderen1, Stuart B Hooper2, Jacco K Kroese1
1Division of Neonatology, Department of Pediatrics, Leiden University Medical Center, Leiden, The Netherlands.
Objective:
To examine the effect of time after birth on heart rate (HR) measured by pulse oximetry (PO) (HRPO) and electrocardiography (ECG) (HRECG).
Study Design:
HRECG and HRPO (collected at maximum sensitivity) were assessed in 53 term and preterm infants at birth. ECG electrodes and a PO sensor were attached as soon as possible and HRECG and HRPO were compared every 30 seconds from 1-10 minutes after birth. Data were compared using a Wilkinson signed-rank test. Clinical relevance (eg, HR <100 beats per minute [bpm] was tested using a McNemar test).
Results:
Seven hundred fifty-five data pairs were analyzed. Median (IQR) gestational age was 37 (31-39) weeks. Mean (SD) starting time of PO and ECG data collection was 99 (33) vs 82 (26) seconds after birth (P = .001). In the first 2 minutes after birth, HRPO was significantly lower compared with HRECG (94 (67-144) vs 150 (91-153) bpm at 60 seconds (P < .05), 81 (60-109) vs 148 (83-170) bpm at 90 seconds (P < .001) and 83 (67-145) vs 158 (119-176) at 120 seconds (P < .001). A HR <100 bpm was more frequently observed with a PO than ECG in the first 2 minutes (64% vs 27% at 60 seconds (P = .05), 56% vs 26% at 90 seconds (P < .05) and 53% vs 21% at 120 seconds (P < .05). HR by ECG was verified by ultrasound for outflow from a subset of infants.
Conclusions:
In infants at birth, HRPO is significantly lower compared with ECG with clinically important differences in the first minutes.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Measurement of Blood Pressure