Related Experiment Video
Updated: Apr 20, 2026

Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Thermal neutron response of a boron-coated GEM detector via GEANT4 Monte Carlo code
1Division of International Studies, College of Global Integrated Studies, Konkuk University, Seoul 143-701, South Korea; IAP, High Energy Physics Lab, Department of Physics, Konkuk University, Seoul 143-701, South Korea.
Abstract:
In this work, we report the design configuration and the performance of the hybrid Gas Electron Multiplier (GEM) detector. In order to make the detector sensitive to thermal neutrons, the forward electrode of the GEM has been coated with the enriched boron-10 material, which works as a neutron converter. A total of 5×5cm2 configuration of GEM has been used for thermal neutron studies. The response of the detector has been estimated via using GEANT4 MC code with two different physics lists. Using the QGSP_BIC_HP physics list, the neutron detection efficiency was determined to be about 3%, while with QGSP_BERT_HP physics list the efficiency was around 2.5%, at the incident thermal neutron energies of 25meV. The higher response of the detector proves that GEM-coated with boron converter improves the efficiency for thermal neutrons detection.
More Related Videos
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
07:51Measurements of Soil Carbon by Neutron-Gamma Analysis in Static and Scanning Modes
Published on: August 24, 2017