Preparation and Pathogen Inactivation of Double Dose Buffy Coat Platelet Products using the INTERCEPT Blood System
Published on: December 7, 2012
Practical issues that should be considered when planning the implementation of pathogen reduction technology for
Teresa Jimenez-Marco1, Catalina Mercant2, Esperanza Lliteras2
1Fundació Banc de Sang i Teixits de les Illes Balears, Majorca, Spain; Institut Universitari d'Investigació en Ciències de la Salut (IUNICS), Spain.
Programming higher platelet yields for plateletpheresis donations can compensate for losses during pathogen reduction technology (PRT) processing. Both Amicus and Trima apheresis devices are suitable for this, with Amicus showing better efficiency and collection rates.
Area of Science:
- Blood banking and transfusion medicine
- Apheresis technology
- Pathogen reduction technology
Background:
- Pathogen reduction technology (PRT) enhances blood safety by inactivating viruses, bacteria, and parasites.
- Diluting platelet (PLT) concentrates in platelet additive solution (PAS) is necessary for PRT application.
- This dilution presents challenges, including increased PLT target yields, extended donation times, and PLT aggregate formation.
Purpose of the Study:
- To evaluate the feasibility of programming higher target PLT yields for plateletpheresis donations to offset PRT-related PLT losses.
- To compare the performance of Amicus 3.11 and Trima 5.22 apheresis separators in collecting hyper-concentrated PLTs for PRT.
Main Methods:
- A paired study involving 196 apheresis procedures on 98 donors was conducted using Amicus 3.11 and Trima 5.22 separators.
- The study compared collection efficiency (CE), collection rate (CR), donation time, and the incidence of PLT aggregates.
- Analysis included a subgroup of donors with higher pre-procedure PLT counts.
Main Results:
- The Amicus 3.11 demonstrated superior CE (78.02% vs. 69.63%) and CR (8.3 vs. 7.00), with a shorter donation time (56.92 vs. 62.60 minutes) compared to the Trima 5.22.
- Similar productivity was observed between the devices for donors with higher pre-procedure PLT counts.
- A significantly higher percentage of PLT aggregates was detected with the Trima (8.62%) versus the Amicus (3.88%).
Conclusions:
- Both apheresis separators are suitable for collecting hyper-concentrated PLTs for PRT with PAS dilution, without significantly extending donation times.
- While PLT aggregation can occur, it largely resolves by day 1 post-collection.
- Further research is warranted to determine the clinical implications of post-apheresis PLT aggregation.