Use of a megavoltage electronic portal imaging device to identify prosthetic materials
Vaughan Moutrie1, Tanya Kairn, Anatoly Rosenfeld
1Department of Radiation Oncology, Princess Alexandra Hospital, 199 Ipswich Rd, Woolloongabba, QLD, 4102, Australia, vaughan.moutrie@health.qld.gov.au.
Abstract:
To achieve accurate dose calculations in radiation therapy the electron density of patient tissues must be known. This information is ordinarily gained from a computed tomography (CT) image that has been calibrated to allow relative electron density (RED) to be determined from CT number. When high density objects such as metallic prostheses are involved, direct use of the CT data can become problematic due to the artefacts introduced by high attenuation of the beam. This requires manual correction of the density values, however the properties of the implanted prosthetic are not always known. A method is introduced where the RED of such an object can be determined using the treatment beam of a linear accelerator with an electronic portal imaging device. The technique was tested using a metallic hip replacement that was placed within a container of water. Compared to the theoretical RED of 6.8 for cobalt-chromium alloy, these measurements calculated a value of 6.4 ± 0.7. This would allow the distinction of an implant as Co-Cr or steel, which have similar RED, or titanium, which is much less dense with an RED of 3.7.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
09:10Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021