Related Experiment Video
Updated: Apr 18, 2026

Assessment of Boron Doped Diamond Electrode Quality and Application to In Situ Modification of Local pH by Water Electrolysis
Published on: January 6, 2016
Monte Carlo modelling the dosimetric effects of electrode material on diamond detectors
Florentina Baluti1, Hossain M Deloar, Stuart P Lansley
1Department of Medical Physics and Bioengineering, Christchurch Hospital, Christchurch, New Zealand, florentina.baluti@cdhb.health.nz.
Abstract:
Diamond detectors for radiation dosimetry were modelled using the EGSnrc Monte Carlo code to investigate the influence of electrode material and detector orientation on the absorbed dose. The small dimensions of the electrode/diamond/electrode detector structure required very thin voxels and the use of non-standard DOSXYZnrc Monte Carlo model parameters. The interface phenomena was investigated by simulating a 6 MV beam and detectors with different electrode materials, namely Al, Ag, Cu and Au, with thickens of 0.1 µm for the electrodes and 0.1 mm for the diamond, in both perpendicular and parallel detector orientation with regards to the incident beam. The smallest perturbations were observed for the parallel detector orientation and Al electrodes (Z = 13). In summary, EGSnrc Monte Carlo code is well suited for modelling small detector geometries. The Monte Carlo model developed is a useful tool to investigate the dosimetric effects caused by different electrode materials. To minimise perturbations cause by the detector electrodes, it is recommended that the electrodes should be made from a low-atomic number material and placed parallel to the beam direction.
Related Concept Videos
Modeling of Diode Forward Characteristics
Modeling of Diode Reverse Characteristics
When a reverse voltage applied to a Zener diode exceeds its breakdown voltage, the diode enters the breakdown region. At this point, the...
Determination of Crystal Structures
Biological Effects of Radiation
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...