Interference of apex locator, pulp tester and diathermy on pacemaker function
Narayanan Sriman1, V Prabhakar1, J S Bhuvaneswaran2
1Department of Conservative Dentistry and Endodontics, Sri Ramakrishna Dental College and Hospital, Coimbatore, India.
Three electronic apex locators (EALs) and the electric pulp tester (EPT) are safe for pacemaker function. However, diathermy significantly interferes with pacemaker function, causing inhibition.
Area of Science:
- Endodontics
- Biomedical Engineering
- Cardiology
Background:
- Electronic apex locators (EALs) and electric pulp testers (EPT) are common dental tools.
- Pacemakers are critical for patients with cardiac conditions.
- Potential electromagnetic interference (EMI) between dental devices and pacemakers requires investigation.
Purpose of the Study:
- To assess the in vitro effects of three EALs, an EPT, and diathermy on pacemaker function.
- To determine the safety of using these dental devices in patients with pacemakers.
Main Methods:
- Three EALs (Root ZX, Propex, Mini Apex locator), an EPT, and diathermy were tested.
- A pacemaker (Medtronic Kappa KVDD901) lead was directly connected to the devices.
- Pacemaker activity was monitored via telemetry and electrocardiogram (EGM) for interference.
Main Results:
- All tested EALs showed no interference with pacemaker function.
- The EPT exhibited minor background noise on EGM but did not affect pacing.
- Diathermy caused irregular pacing and complete pacemaker inhibition.
Conclusions:
- Electronic apex locators are safe for use in patients with pacemakers.
- Electric pulp testers may cause minor noise but do not compromise pacemaker function.
- Diathermy poses a significant risk and should be avoided in pacemaker patients.
More Related Videos
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Holter Monitor: 24-Hour Monitoring
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection