A Novel Computerized Portable Pupillometer Detects and Quantifies Relative Afferent Pupillary Defects
Liza M Cohen1, Michael A Rosenberg1, Angelo P Tanna1
1a Department of Ophthalmology , Northwestern University Feinberg School of Medicine , Chicago , IL , USA.
Current Eye Research
|February 7, 2015
Summary
A new portable pupillometer accurately detects and quantifies relative afferent pupillary defects (RAPDs). This device shows high sensitivity and specificity, aiding in screening patients with vision loss.
Area of Science:
- Ophthalmology
- Neuroscience
- Medical Devices
Background:
- Relative afferent pupillary defects (RAPDs) are crucial indicators of optic nerve and retinal dysfunction.
- Accurate and objective quantification of RAPDs is essential for diagnosing and monitoring vision loss.
- Current methods for RAPD assessment can be subjective and operator-dependent.
Purpose of the Study:
- To validate a novel computerized portable pupillometer for detecting and quantifying RAPDs.
- To assess the device's correlation with expert clinical grading of RAPDs.
- To determine the sensitivity and specificity of the pupillometer in identifying clinically significant RAPDs.
Main Methods:
- Pupillary response curves were recorded in patients with clinically graded RAPDs (n=32) and normal subjects (n=31) with simulated RAPDs.
- Pupillary constriction amplitude (CA), velocity (CV), and onset latency (COL) were used to calculate RAPDs.
- Inter-visit variability was assessed in 14 normal subjects.
Main Results:
- The novel pupillometer demonstrated a strong correlation with expert clinical grading of RAPDs (r=0.81, p<0.0001).
- The device distinguished patients with RAPDs (≥0.5 LU) from normal subjects with 91% accuracy.
- Calculations using CA and CV showed stronger correlation with clinical grading than COL.
Conclusions:
- The computerized portable pupillometer is a reliable tool for objective RAPD assessment.
- The device exhibits high sensitivity and specificity for detecting clinically significant RAPDs.
- This pupillometer has potential as a valuable clinical tool for screening patients with vision loss.
Keywords:
Diagnostic testspupilpupillometryrelative afferent pupillary defectswinging flashlight test