[Validation during exercise of a new device for cardiac output measurement using pulse wave transit time (comparison
Objectives:
EsCCO is a novel non-invasive continuous cardiac output monitoring system based on pulse wave transit time already validated at rest. The aim of our study was to compare cardiac output measurements obtained simultaneously by EsCCO(®) (Q˙cOP) and impedance cardiography (Physioflow(®) ; Q˙cIMP), in healthy subjects.
Patients And Methods:
Eight healthy subjects (age: 31±9 years, weight: 76±10kg, height: 179±5cm) realized two exercise tests: an incremental ergocycle test performed until exertion (Pmax=269±48W) and a constant load exercise (P=163±27W). Comparison between measurements (Q˙cOP versus Q˙cIMP) obtained during the first test allowed to evaluate the accuracy of the device. Reliability was determined on three repeated measures during the second test, realized at ventilatory threshold.
Results:
Correlation coefficient between both methods is 0.88 (P<0.01). Mean difference is 0.04±1.49L/min (95 % limits of agreement: +2.94 to -3.00L/min) and only 3/74 measures are not included between the limits of agreement. At high intensity and for cardiac output over than 15 L/min, Q˙cOP signal is lost in almost half the time. Concerning reliability, reproducibility coefficient is 0.87 (P<0.05), only 1.8 % of this variability is due to the method.
Conclusion:
EsCCO(®) measurements are accurate, reliable and allow a good estimation of cardiac output on healthy subjects. The signal lost observed for high cardiac output levels (>15L/min) can limit its utilization during very high intensity exercise.
Related Concept Videos
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: