An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Biomechanical evaluation of resistance to insertion torque of different implant systems and insertion driver types
Hugo Nary Filho1, José Luis Calvo Guirado, Mariza Akemi Matsumoto
1*Oral and Maxillofacial Surgeon and Coordinator, Advanced Implantology Course, P-I Brånemark Institute, São Paulo, Brazil; Coordinator, Graduate Course in Implantodontics, Universidade Sagrado Coração (USC), São Paulo, Brazil. †Professor, Clinica Odontológica Integrada de Adultos, Murcia, Spain; Senior Lecturer, General Dentistry, Faculty of Medicine and Dentistry, University of Murcia, Murcia, Spain. ‡Assistant Professor, Department of Oral and Maxillofacial Surgery, Universidade Sagrado Coração (USC), São Paulo, Brazil. §Fellow, Master Course in Implantodontics, Universidade Sagrado Coração (USC), São Paulo, Brazil.
Purpose:
The aim of this study was to quantify the resistance to torque of different implant systems and their connection devices using in vitro torsion tests.
Materials And Methods:
Three internal connection systems, 1 conventional system with internal torque and 1 conventional system with a mounting device used as a control group were tested on 5 groups.
Results:
Rupture torque (in newton meter): Biomet 3i Certain group 4 showed a statistically significant higher average (2.65 N·m), followed in order by Biomet 3i Osseotite group 5 (2.18 N·m), Bonelike group 2 (1.80 N·m). Angle deformation/rupture: all groups obtained similar values, without significant differences. Elastic limit (in newton meter): Bonelike group 2 (1.06 N·m) showed similar behavior to group 1 (1.39 N·m) (Nobel Biocare), without significant differences, whereas Bonelike group 3 showed a significantly lower value (0.93 N·m). Maximum torque (in newton meter): Biomet 3i Certain group 4 showed significantly higher values in relation to other groups (2.80 N·m).
Conclusions:
The greater contact area the system is built on, the greater resistance against insertion torque, as internal hexagon implants with a greater contact area and external hexagon implants using a mounting device showed higher resistance to insertion torque.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Screw: Problem Solving
To evaluate the...
Residual Stresses in Circular Shafts
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Bearings: Problem Solving
Net Torque Calculations