Voltammetric Determination of Codeine on Glassy Carbon Electrode Modified with Nafion/MWCNTs
Robert Piech1, Martyna Rumin1, Beata Paczosa-Bator1
1Faculty of Material Science and Ceramics, AGH University of Science and Technology, Mickiewicza 30, 30059 Cracow, Poland.
Abstract:
A glassy carbon electrode modified with a Nafion/MWCNTs composite is shown to enable the determination of codeine using differential pulse voltammetry in phosphate buffer of pH 3.0. At a preconcentration time of 15 s, the calibration graph is linear in the 0.5 µM (0.15 mg·L(-1)) to 15 µM (4.5 mg·L(-1)) concentration range with a correlation coefficient of 0.998. The detection limit at a preconcentration time of 120 s is as low as 4.5 μg·L(-1). The repeatability of the method at a 0.6 μg·L(-1) concentration level, expressed as the RSD, is 3.7% (for n = 5). The method was successfully applied and validated by analyzing codeine in drug, human plasma, and urine samples.
More Related Videos
09:17Reductive Electropolymerization of a Vinyl-containing Poly-pyridyl Complex on Glassy Carbon and Fluorine-doped Tin Oxide Electrodes
Published on: January 30, 2015
12:20Fabrication of Carbon Nanotube High-Frequency Nanoelectronic Biosensor for Sensing in High Ionic Strength Solutions
Published on: July 22, 2013
Related Concept Videos
Potentiometry: Membrane Electrodes
Voltammetric Techniques: Cyclic Voltammetry
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...