One-Year Feasibility Study of Replenish MicroPump for Intravitreal Drug Delivery: A Pilot Study
Juan-Carlos Gutiérrez-Hernández1, Sean Caffey2, Walid Abdallah3
1Doheny Eye Institute, Los Angeles, CA ; Centro de Retina Médica y Quirúrgica, SC, Zapopan, Jalisco, Mexico.
Translational Vision Science & Technology
|March 17, 2015
Summary
This study demonstrates the successful surgical implantation and safety of the Replenish Posterior MicroPump for intravitreal drug delivery in Beagle dogs. The device showed good biocompatibility, supporting its use in future ocular automated drug delivery systems.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Drug Delivery Systems
Background:
- Intravitreal drug delivery is crucial for treating posterior segment ocular diseases.
- Novel drug delivery systems are needed to improve therapeutic efficacy and patient compliance.
- Micro-scale devices offer potential for sustained and localized drug release.
Purpose of the Study:
- To assess the feasibility of surgically implanting the Replenish Posterior MicroPump for intravitreal drug delivery.
- To collect safety and biocompatibility data of the micro drug pump in a Beagle dog model for up to one year.
- To evaluate the bioelectronic performance and tissue response to the implanted device.
Main Methods:
- Thirteen Beagle dogs were divided into an experimental group (n=11) receiving MicroPump implantation and a control group (n=2) undergoing sham surgery.
- The MicroPump was implanted via pars plana, with the catheter secured at the sclerotomy site.
- Ophthalmic examinations, imaging, and histopathology were performed at 1, 3, and 12 months post-surgery.
Main Results:
- No clinically significant pathological changes were observed in implanted or control eyes beyond expected surgical alterations.
- The operated eyes exhibited minimal inflammatory reactions and no significant tissue ingrowth at the sclerotomy site.
- The Replenish Posterior MicroPump demonstrated a favorable safety profile in the Beagle dog model.
Conclusions:
- The Replenish Posterior MicroPump can be successfully implanted surgically with a good safety profile in this animal model.
- The study supports the biocompatibility of the Replenish MicroPump for ocular automated drug delivery.
- Further testing in larger animal models is warranted to advance the use of these devices.