Aspects of three dimensional reconstruction for a multi ring positron tomograph
D Townsend1, T Spinks, T Jones
1MRC Cyclotron Unit, Hammersmith Hospital, London, UK.
European Journal of Nuclear Medicine
|January 1, 1989
Summary
Removing septa in positron emission tomography (PET) scanners improves signal-to-noise ratio in the center but increases scatter and randoms. This enables 3D reconstruction for enhanced imaging within a limited axial range.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Positron Emission Tomography (PET)
Background:
- Inter-plane septa in multi-ring PET scanners reduce random and scattered coincidences.
- Septa typically limit coincidence detection to adjacent detector rings, excluding lines of response between non-adjacent rings.
Purpose of the Study:
- To evaluate the impact of removing septa on image quality and signal-to-noise ratio (SNR) in a clinical PET scanner.
- To assess the feasibility of 3D reconstruction with septa removed for enhanced data acquisition.
Main Methods:
- Utilized an 8-ring, 15-slice ECAT 931/08-12 scanner with a configuration option to remove septa.
- Acquired full 64 sinograms with septa removed, using a rotating pin source for detector normalization and transmission data.
- Reconstructed phantom studies using a fully three-dimensional (3D) algorithm and compared results with septa-in data.
Main Results:
- Signal-to-noise ratio (SNR) improved by a factor of 2.8 in the center of a uniform 20 cm diameter cylinder.
- SNR decreased in regions distant from the center axially due to increased scatter and random coincidences.
- An overall improvement in SNR was observed within 6 cm of the tomograph's 10 cm axial length.
Conclusions:
- Removing septa in PET imaging can significantly enhance SNR in central regions, particularly when using 3D reconstruction.
- The benefits of septa removal are spatially dependent, with potential trade-offs in axial regions due to increased scatter and randoms.
- Optimized 3D reconstruction strategies are crucial for maximizing the advantages of septa-less PET acquisition.