A Protocol for Analyzing Hepatitis C Virus Replication
Published on: June 26, 2014
Extraction of HCV-RNA from Plasma Samples: Development towards Semiautomation
Imran Amin1, Tania Jabbar2, Fawad Niazi1
1PCR Laboratory, Punjab Institute of Nuclear Medicine (PINUM), Faisalabad 2019, Pakistan.
Abstract:
A semiautomated extraction protocol of HCV-RNA using Favorgen RNA extraction kit has been developed. The kit provided protocol was modified by replacing manual spin steps with vacuum filtration. The assay performance was evaluated by real-time qPCR based on Taqman technology. Assay linearity was confirmed with the serial dilutions of RTA (Turkey) containing 1 × (10(6), 10(5), 10(4), and 10(3)) IU mL(-1). Comparison of test results obtained by two extraction methods showed a good correlation (r = 0.95, n = 30) with detection limit of 10(2) IU mL(-1). The semiautomated vacuum filtration based protocol demonstrated high throughput: 35 minutes for the extraction of a batch of 30 samples (150 µL each) with reduced labor, time, waste, and cost. Performance characteristics of semiautomated system make it suitable for use in diagnostic purpose and viral load determinations.