Related Experiment Video
Updated: Apr 14, 2026

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
[Liquid chromatography for determination of disulfiramin workplace atmosphere]
Sheng Fu1, Zhihui Han1, Keping Yu1
1Hunan Prevention And Treatment Institute For Occupational Disease, Changsha 410007, China.
Objective:
To establish a method for determination of disulfiram in the workplace atmosphere by liquid chromatography.
Methods:
Sampling with glass fiber filter, eluting with methanol, separating with C18 column, and determination with liquid chromatography.
Results:
The bearing capacity of glass fiber filter exceeded 3.45 mg per piece. The elution efficiency was 97.8%∼101.0% The relative standard deviation varied from 1.09% to 1.44%. The limit of detection was 0.1 µg/ml. The minimum detectable concentration was 0.011 mg/m³ (with sampled air volume of 45 L).
Conclusion:
The method has high selectivity, accuracy, and precision and strong applicability.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-II