Semi-automated echocardiographic quantification of right ventricular size and function
Diego Medvedofsky1, Karima Addetia, Jamie Hamilton
1Section of Cardiology, Department of Medicine, University of Chicago Medical Center, 5841 South Maryland Avenue, MC5084, Chicago, IL, 60637, USA.
Summary
A new semi-automated software provides fast and accurate measurements of right ventricular (RV) size and function using speckle tracking. This method significantly reduces analysis time compared to traditional manual techniques.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Assessing right ventricular (RV) size and function is crucial but challenging with conventional echocardiography.
- Current quantitative methods are time-consuming, often leading to reliance on qualitative assessments in clinical practice.
Purpose of the Study:
- To evaluate a novel semi-automated speckle tracking software for measuring RV size and function.
- To compare the accuracy and efficiency of this automated approach against conventional manual echocardiographic measurements.
Main Methods:
- Transthoracic echocardiograms from 149 patients with diverse RV conditions were analyzed.
- A semi-automated software (Epsilon Imaging) was used, requiring single-frame endocardial boundary tracing.
- Key RV parameters including areas, dimensions, fractional area change, and systolic excursion were measured and compared.
Main Results:
- The semi-automated method demonstrated high feasibility (90%) and significantly reduced analysis time (30 seconds vs. 4 minutes).
- Measurements showed good agreement with manual techniques, with correlation coefficients (r) ranging from 0.70 to 0.95.
- Biases were within 2-22% of mean values, comparable to the inherent variability of manual measurements.
Conclusions:
- The semi-automated speckle tracking technique is a feasible and rapid tool for quantitative RV assessment.
- This approach provides reliable RV size and function parameters comparable to conventional methods, enhancing clinical efficiency.