Value of C-Arm Computed Tomography to Evaluate Stent Deployment During Femoro-Popliteal Revascularization
Gerald Gahide1, Sofiane Hadjadj2, Eric Therasse2,3,4
1Centre de Recherche Etienne LeBel, Centre Hospitalier de l'Université de Sherbrooke, Sherbrooke, QC, Canada.
Purpose:
To compare the accuracy of C-arm computed tomography (CT) and digital subtraction angiography (DSA) in detecting incomplete stent expansion (ISE) after superficial femoral artery (SFA) stenting using intravascular ultrasound (IVUS) as a gold standard.
Materials:
Fifty patients with symptomatic SFA occlusive disease requiring angioplasty were prospectively included. Once technical success (<30 % residual stenosis) was obtained on post-procedural DSA, C-arm CT and IVUS were acquired. DSA and C-arm CT examinations were reviewed by 2 investigators and correlated with IVUS. C-arm CT image quality was rated on a four-point scale. Doppler ultrasound was performed at 1-year follow-up.
Results:
The ankle-brachial index was 0.69 ± 0.10 and 0.99 ± 0.40, respectively, pre- and post-procedure. C-arm CT imaging quality was rated as good or excellent in 80%. In-stent minimal luminal diameter (MLD) was evaluated at 4.71 ± 0.7 mm on DSA, 3.39 ± 0.6 mm on IVUS, and 3.12 ± 0.9 mm on C-arm CT. Compared to IVUS, DSA demonstrated an overestimation of MLD (p = 0.0001), an underestimation of ISE (DSA = 18.8% ± 7.6; IVUS = 29.8% ± 9) (p < 0.0001), and a poor inter-technique intra-class correlation coefficient (ICC = 0.24). No difference was observed between IVUS and C-arm CT in ISE as calculated by diameter (29.8 ± 9 vs. 28.2 ± 12.5%, p = 0.5) and area (30.2 ± 8.4 vs. 33.3 ± 9.5%, p = 0.2). Inter-technique ICC between C-arm CT and IVUS was 0.72 [95%CI 0.49; 0.85] for MLA measurements. The inter-observer ICC for MLD and MLA measurements on C-arm CT were, respectively, estimated at 0.75 [95% CI 0.40, 0.89] and 0.77 [95% CI 0.43, 0.90)].
Conclusions:
C-arm CT presents a better correlation with IVUS than DSA to determine lumen diameter and ISE immediately after percutaneous revascularization.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography