On-orbit characterization of the VIIRS solar diffuser and solar diffuser screen.
Applied Optics
|May 14, 2015
Summary
On-board calibration of the Visible Infrared Imaging Radiometer Suite (VIIRS) was improved using bidirectional reflectance factors (BRF) and vignetting functions (VF). This enhances accuracy for satellite ocean color remote sensing products.
Area of Science:
- Remote Sensing
- Instrument Calibration
- Optical Physics
Background:
- On-board calibration is crucial for accurate satellite data, especially for ocean color products.
- The Visible Infrared Imaging Radiometer Suite (VIIRS) relies on solar diffuser (SD) and its screen (SDS) for calibration.
- Degradation of the SD impacts data quality over time.
Purpose of the Study:
- To refine the on-board calibration of the VIIRS instrument by analyzing the bidirectional reflectance factors (BRF) of the solar diffuser (SD) and the vignetting function (VF) of the SD screen (SDS).
- To improve the accuracy and consistency of satellite ocean color remote sensing products through enhanced instrument characterization.
- To determine the absolute form of the BRF-VF product and the SDS VF using spacecraft yaw maneuver data.
Main Methods:
- Conducted 14 spacecraft yaw maneuvers for the Suomi National Polar-Orbiting Partnership satellite to gather data on VIIRS instrument response to solar angles.
- Combined yaw maneuver data with pre-launch measurements of SDS VF and SD BRF.
- Determined the absolute BRF-VF product and SDS VF for reflective solar bands (RSB) and the solar diffuser stability monitor (SDSM) detectors.
Main Results:
- The BRF-VF product for RSB was found to be independent of detector, gain status, and half-angle mirror side.
- Derived VFs showed agreement with pre-launch measurements and primarily depended on solar angles.
- Derived calibration coefficients (F-factors) demonstrated a significantly improved pattern, with residual seasonal fluctuations corrected by a solar azimuth angle-dependent function.
Conclusions:
- The refined calibration approach using BRF and VF significantly improves the accuracy of VIIRS instrument characterization.
- The results directly benefit satellite ocean color remote sensing by providing more reliable data.
- The study establishes a robust method for on-orbit calibration, reducing uncertainties in long-term data records.
More Related Videos
09:19In Situ Monitoring of the Accelerated Performance Degradation of Solar Cells and Modules: A Case Study for CuIn,GaSe2 Solar Cells
Published on: October 3, 2018
8.9K
07:32Printing Fabrication of Bulk Heterojunction Solar Cells and In Situ Morphology Characterization
Published on: January 29, 2017
11.7K
Related Concept Videos
UV–Vis Spectrometers
4.7K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
4.7K
IR Spectrometers
3.5K
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
3.5K
UV–Vis Spectrum
3.4K
When light passes through a substance, a portion of the light is absorbed while the remaining light is reflected or transmitted. If the molecule absorbs light between the wavelengths of 180–400 nm range, the UV spectrum is obtained, and if it absorbs light in the 400–780 nm wavelength range, the visible spectrum is obtained.
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
3.4K