Neonatal nasal intermittent positive pressure ventilation efficacy and lung pressure transmission
1Division of Neonatology, Department of Paediatrics, McMaster Children's Hospital, McMaster University, Hamilton, ON, Canada.
Objective:
The objective of this study was to evaluate carbon dioxide (CO2) clearance, delivered pressures and tidal volume (VT) during neonatal nasal intermittent positive pressure ventilation (NIPPV) with two commonly used interfaces.
Study Design:
A neonatal lung model, with either short binasal prongs (SBP) or a small caliber nasal cannula (RAM) interface, was tested over a range of clinically relevant settings. A fixed amount of CO2 was infused and the fraction remaining in the lung 100 s postinfusion was measured. Pressure transmission to the lung and VT was measured at the level of the trachea.
Result:
CO2 elimination was directly proportional to the inspiratory pressure during NIPPV. At peak pressures of 22 to 34 cm H2O, CO2 clearance was greater (P<0.001) with SBP as compared with RAM. Relative to the set ventilator parameters, a substantial pressure dampening effect was documented at the lung level, which was significantly lower with RAM when compared with SBP (2.8% (0.2) versus 11.9% (1.5), P<0.0001). CO2 elimination was dependent on VT and effective despite only a small fraction of physiological VT (maximum delivered VT%: SBP 15.5 (0.7) versus RAM 6.1 (1.4), P<0.0001).
Conclusion:
NIPPV promotes CO2 elimination even at low transmitted airway pressures, but less effective with RAM as compared with SBP. CO2 elimination despite small VT suggests that NIPPV may depend on a non-conventional gas-exchange mechanism.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Suctioning the Nasopharyngeal Airway
Equipment Required