Related Experiment Video
Updated: Apr 9, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
A New Design of a Lead-Acrylic Shield for Staff Dose Reduction in Radial and Femoral Access Coronary Catheterization
H Eder1, M C Seidenbusch2, M Treitl2
1Department of Radiation Protection, formerly: Bavarian Office for Occupational Health und Safety.
Purpose:
Today's standard radiation protection during coronary angiography and percutaneous coronary interventions is the combined use of lead acrylic shields and table-mounted lower body protection. Ambient dose measurements, however, have shown that these protection devices need improvement.
Materials And Methods:
Using an anthropomorphic physical phantom, various scenarios were investigated with respect to personnel exposure: a) enlarging the shield b) adding a flexible protective curtain to the bottom side of the shield, and c) application of radioprotective patient drapes. For visualization of the dose reduction effect, Monte Carlo simulations were performed.
Results:
The flexible curtain in contact with the patient's body reduces the ambient dose rate at the operator's position by up to (87.5% ± 7.1) compared to the situation with the bare shield. The use of both the flexible curtain and the patient drape reduces the ambient dose rate by up to (90.8% ± 7). Similar results were achieved for the assisting personnel when they were positioned next to the operator. In addition, the enlarged shield provides better protection of the head region of tall operators.
Conclusion:
Adding a flexible protective curtain to the bottom side of the shield can protect operators from high doses, especially for body parts which are not protected by lead aprons, e.g. head, and eye lenses. This may be important with respect to lower dose limits for eye lenses in future. The protective effect in real-life working conditions is still being evaluated in an ongoing clinical study.
Key Points:
Lead acrylic shields need improvement for a better protection of head and eye lenses. An additional flexible lead curtain at the bottom of the shield can considerably reduce the operator dose. Using the additional lead curtain, lighter protection clothing can be worn. Special eye protection may be no longer needed in most applications.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
05:31Transradial Access Chemoembolization for Hepatocellular Carcinoma Patients
Published on: September 20, 2020
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization