Design and implementation of a "cheese" phantom-based Tomotherapy TLD dose intercomparison.
Hans Schiefer1, Konrad Buchauer2, Simon Heinze2
1Department of Radiation Oncology, Medical Physics Group, Kantonsspital St. Gallen, 9007, St. Gallen, Switzerland. johann.schiefer@kssg.ch.
The "cheese" phantom is more suitable for TomoTherapy quality assurance dose checks than the Perspex phantom. This finding will guide future Swiss Society of Radiobiology and Medical Physics dosimetry intercomparisons for TomoTherapy machines.
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- TomoTherapy machines require specific quality assurance (QA) protocols due to their unique beam delivery.
- External dose intercomparisons are crucial for ensuring accurate radiation delivery.
- The 2014 Swiss Society of Radiobiology and Medical Physics (SSRMP) dosimetry intercomparison aimed to evaluate different phantom setups for TomoTherapy QA.
Purpose of the Study:
- To compare the suitability of two different phantom setups for external dose intercomparisons in TomoTherapy.
- To assess the accuracy of dose measurements using the standard calibration plans for TomoHelical and TomoDirect techniques.
Main Methods:
- Two phantoms were used: a cylindrical Perspex phantom and the manufacturer-provided "cheese" phantom.
- Standard calibration plans for TomoHelical and TomoDirect irradiation were applied.
- Thermoluminescent dosimetry (TLD) was used to measure the dose, and the ratio of measured dose (Dm) to calculated dose (Dc) was assessed for 20 TomoTherapy machines.
Main Results:
- The standard deviation of Dm/Dc ratios for the "cheese" phantom was 1.9% for TomoHelical and 1.2% for TomoDirect.
- For the Perspex phantom, the standard deviations were 2.8% for TomoHelical and 1.8% for TomoDirect.
- The "cheese" phantom demonstrated lower variability in dose measurements.
Conclusions:
- The "cheese" phantom setup, without individual planning, is more suitable for TomoTherapy dose checks compared to the Perspex phantom setup.
- Future SSRMP dosimetry intercomparisons for TomoTherapy will utilize the "cheese" phantom setup.