Related Experiment Video
Updated: Apr 9, 2026

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Pulmonary Vein Isolation Using the First-Generation Cryoballoon Technique in Chinese Patients
Gong-Bu Zhou1, Xiao-Gang Guo1, X U Liu1
1State Key Laboratory of Cardiovascular Disease, Arrhythmia Center, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, People's Republic of China.
Background:
Cryoballoon ablation is an alternative for the treatment of atrial fibrillation (AF). We assess the relationship among pulmonary vein (PV) electrophysiology, balloon temperature, and persistent PV isolation (PVI), and report procedural properties, safety, and efficacy of this technique in a Chinese center for the first time.
Methods:
A total of 70 patients with paroxysmal (n = 57) and persistent (n = 13) AF were consecutively enrolled. PVI was performed with cryoballoon catheter. Real-time recording of dissociation of PV potentials was attempted using a circumferential mapping catheter.
Results:
Successful PVI was achieved in 274 of 282 (97.2%) PVs. Procedural duration was 115.2 ± 24.8 minutes and fluoroscopy time was 29.6 ± 8.9 minutes. Real-time PV potential recording was achieved in 232 (84.3%) PVs. Regarding time to isolation, a cut-off value of less than 60 seconds was predictive of persistent PVI with a sensitivity of 0.76 and specificity of 0.82 (area under curve = 0.835; P < 0.0001). The nadir balloon temperature was significantly lower in PVs without early reconduction (-46.3 ± 0.5°C vs -40.2 ± 1.3°C, P < 0.0001). However, there was an insignificant trend that the balloon temperature at isolation was higher in PVs without reconduction (-33.9 ± 0.7°C vs -36.4 ± 2.0°C, P = 0.14). The overall complication rate was 5.7%. After a median follow-up of 6.5 (range 3.2-14.9) months, 76% of patients were free of AF recurrence with a blanking period of 3 months.
Conclusions:
Real-time PV potentials can be recorded in most PVs. The time to isolation is predictive of the persistency of PVI. The nadir balloon temperature, not the balloon temperature at isolation, is significantly lower in PVs without reconduction.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020