Related Experiment Video
Updated: Apr 7, 2026

Manufacture of a Multi-Purpose Low-Cost Animal Bench-Model for Teaching Tracheostomy
Published on: May 18, 2019
Experimental Assessment and Future Applications of the Shikani Tracheostomy Speaking Valve
Purpose:
Tracheostomy speaking valve use may increase airflow resistance and work of breathing. It remains unclear which valve offers the best performance characteristics. We compared the performance characteristics of the Shikani speaking valve (SSV; unidirectional-flow ball valve) with those of the Passy-Muir valve (PMV; bias-closed flapper valve).
Method:
Airflow resistance was measured for both the SSV and the PMV at 8 flow amplitudes and in 3 orientations (-15°, 0°, +20°) in the bias-open and bias-closed configurations.
Results:
Significantly lower airflow resistance was observed for the SSV (bias open) compared with the PMV at -15° (p < .001), 0° (p < .001), and +20° (p = .006) from the horizon. No significant difference was observed between the PMV and the SSV (bias-closed) configuration at any of the tested angles. A nonsignificant trend toward decreased airflow resistance was observed between the SSV bias-open and bias-closed configurations at each of the angles tested.
Conclusions:
The SSV demonstrated lower airflow resistance compared with the PMV across 8 flow amplitudes in the bias-open configuration at -15°, 0°, and +20° from the horizon. Further investigation is needed to determine the clinical impact of these findings on patient comfort, work of breathing, phonation, and airway protection during swallowing.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...