Ventilator disconnexion alarm failures. The role of ventilator and breathing system accessories
1Shackelton Department of Anaesthetics, Southampton General Hospital.
Abstract:
Pressure-sensitive ventilator disconnexion alarms do not always alarm during disconnexion of a discharging compliance ventilator such as Manley Blease, unless accurately adjusted. High flows during disconnexion result in significant pressure generation caused by outflow resistance of catheter mounts, heat and moisture exchangers, capnometer cuvettes, and angled connectors; this may lead to alarm failure because of incorrectly adjusted pressure alarm limits. The exact position of the disconnexion is critical and if the alarm's pressure sensor is placed in either the inspiratory or expiratory limb of the ventilator it makes no difference to its correct function. Nine different heat and moisture exchanges were compared in the same breathing system. Those with 15-mm male connectors generate the highest pressures on disconnexion (1.1 kPa). It is suggested that the low pressure alarm limit is set only marginally below the peak inspiratory pressure, and that it is readjusted for every patient and after every change in ventilation. Most importantly, the alarm should be shown to be functional by a trial disconnexion at the tracheal tube.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
04:16Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...