Abstract:
A suppressed hydroxide eluent anion chromatograph effluent flows through the outside of a gas-permeable membrane tube while electrogenerated 100-200 μM LiOH flows through the lumen into a second conductivity detector. Undissociated volatile acid eluites (e.g., H2S, HCN, H2CO3, etc., represented as HA) transfer through the membrane and react as OH(-) + HA → A(-) + H2O; the conversion of high-mobility OH(-) to lower mobility A(-) results in a significant negative response for these analytes. With the chromatograph operated at a macroscale (0.3 mL/min) the LiOH flow can be 3-30-fold lower, resulting in corresponding enrichment of the transferred analyte prior to detection. Because there is no mixing of liquids, the detector noise is very low (<0.1 nS/cm), comparable to the principal chromatographic detector. Thus, despite a background of 25-45 μS/cm, limits of detection for sulfide and cyanide are in the submicromolar level, with a linear dynamic range up to 100 μM. Carbonate/bicarbonate can also be sensitively detected. We demonstrate adaptation in a standard commercial system. We also show that Microsoft Excel-based numerical simulations of transport quantitatively predict the observed behavior well.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Ion-Exchange Chromatography
Conductometric Titrations: Strong Acid-Base and Weak Acid-Base Titrations
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...