Retrospective evaluation of exposure index (EI) values from plain radiographs reveals important considerations for
Ursula Mothiram1, Patrick C Brennan1, John Robinson1
1Medical Imaging Optimisation and Perception Group (MIOPeG), Discipline of Medical Radiation Sciences, Faculty of Health Science, University of Sydney Sydney, Australia.
Introduction:
Following X-ray exposure, radiographers receive immediate feedback on detector exposure in the form of the exposure index (EI).
Purpose:
To identify whether radiographers are meeting manufacturer-recommended EI (MREI) ranges for routine chest, abdomen and pelvis X-ray examinations under a variety of conditions and to examine factors affecting the EI.
Methods:
Data on 5000 adult X-ray examinations including the following variables were collected: examination parameters, EI values, patient gender, date of birth, date and time of examination, grid usage and the presence of implant or prosthesis. Descriptive statistics were used to summarize each data set and the Mann-Whitney U test was used to determine significant differences, with P < 0.05 indicating significance for all tests.
Results:
Most examinations demonstrated EI values that were outside the MREI ranges, with significantly higher median EI values recorded for female patient radiographs than those for male patients for all manufacturers, indicating higher detector exposures for all units except for Philips digital radiography (DR), where increased EI values indicate lower exposure (P = 0.01). Median EI values for out of hours radiography were also significantly higher compared with normal working hours for all technologies (P ≤ 0.02). Significantly higher median EI values were demonstrated for Philips DR chest X-rays without as compared to those with the employment of a grid (P = 0.03), while significantly lower median EI values were recorded for Carestream Health computed radiography (CR) chest X-rays when an implant or prosthesis was present (P = 0.02).
Conclusions:
Non-adherence to MREIs has been demonstrated with EI value discrepancies being dependent on patient gender, time/day of exposure, grid usage and the presence of an implant or prosthesis. Retrospective evaluation of EI databases is a valuable tool to assess the need of quality improvement in routine DR.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Biological Effects of Radiation
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...