Related Experiment Video
Updated: Apr 5, 2026

Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
Absorb bioresorbable vascular scaffold: What have we learned after 5 years of clinical experience?
Gianluca Caiazzo1, Ismail Dogu Kilic2, Enrico Fabris3
1Division of Cardiology, Department of Medical and Surgical Sciences, "Magna Graecia" University, Catanzaro, Italy; Cardiovascular National Institute of Health Research Biomedical Research Unit, Royal Brompton & Harefield NHS Foundation Trust, London, United Kingdom.
Abstract:
Bioresorbable scaffolds have the potential to introduce a paradigm shift in interventional cardiology, a true anatomical and functional "vascular restoration" instead of an artificial stiff tube encased by persistent metallic foreign body. Early clinical studies using the first commercially available drug-eluting bioresorbable vascular scaffold (BVS) reported very promising safety and efficacy outcomes, comparable to best-in-class second-generation drug-eluting metal stent. To date, more than 60,000 Absorb BVSs have been implanted with only the interim analysis of one randomized trial (ABSORB II RCT) available. Recent registries have challenged the initial claim that BVS is immune from Scaffold Thrombosis (ST). However, suboptimal device expansion and insufficient intracoronary imaging guidance can explain higher than expected ST, especially in complex lesions. The aim of this review article is to critically evaluate the results of the available Absorb BVS studies and discuss the lessons learned to optimize lesion selection and implantation technique of such devices.