Related Experiment Video
Updated: Apr 4, 2026

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Left atrial appendage closure: First in man with the 4th generation WATCHMAN device
Jakob Ledwoch1, Jennifer Franke1, Maik Gonzaga1
1CardioVascular Center CVC, Frankfurt, Germany.
Objectives:
To evaluate the feasibility and safety of the fourth generation WATCHMAN device.
Background:
The WATCHMAN left atrial appendage (LAA) closure device has been shown to be non-inferior to conventional oral anticoagulation with warfarin for stroke prevention in patients with non-valvular atrial fibrillation. A new (fourth) generation of the WATCHMAN device was designed to facilitate easier delivery and improve safety.
Methods:
We conducted a prospective, non-randomized study of LAA closure with use of the 4(th) generation WATCHMAN device in 36 patients with non-valvular atrial fibrillation. Follow-up was performed at 45 days, 6 months and 12 months following implantation.
Results:
The mean age of the population was 73 ± 6 years and 67% were male. The mean CHADS2 and CHA2 DS2-VASc Score were 2.5 and 4.5, respectively. The device was implanted successfully in 34/36 (94%). No serious hospital complications related to the device or procedure occurred. During 12 month follow-up, three ischemic strokes (8%), one hemorrhagic stroke (3%), one TIA (3%) and three deaths (9%) occurred. Device-associated thrombus formation was detected in one patient (3%) during six-month follow-up and was treated successfully with low molecular weight heparin.
Conclusions:
LAA closure with the 4(th) generation WATCHMAN device is feasible and safe. New technical features may contribute to a better performance of the occluder, particularly regarding the risk of pericardial effusions.