Radiographers' Ability to Detect Low-Contrast Detail in Digital Radiography Systems
Purpose:
To evaluate radiographers' ability to detect low-contrast detail using various digital planar radiographic systems.
Methods:
A low-contrast detail phantom was placed between two 5-cm thick Perspex sheets (Lucite International). Images were obtained using different kilovoltage peak and milliamperage second (mAs) settings with computed radiography (CR), indirect conversion digital radiography (IDR), and direct conversion digital radiography (DR) systems. Six groups of 6 radiographers were asked to score 39 images; each group scored 2 images from each system for a total of 6 images. The seventh group scored only one image from each system for a total of 3 images. The radiographers' results were compared with the results of analyzer software. The inverse image quality factor was used to measure low-contrast detail detectability performance.
Results:
Radiographers performed significantly worse than the computerized software in determining low-contrast detail in planar radiographic images (P < .01). However, a positive correlation (R = 0.558) existed between the 2 sets of scores in terms of low-contrast detail detectability performance.
Discussion:
On average, radiographers were able to detect increased image quality resulting from increased mAs. Radiographers reached results similar to the software regarding whether IDR and DR have better detectability performances than CR. Differences found among individual radiographers were not as significant with DR.
Conclusion:
When radiographers' performance in detecting low-contrast detail was evaluated and compared with that of the software, radiographers exhibited poorer performances. Because radiographers are responsible for image quality optimization, additional training might improve their ability to detect low-contrast detail in DR systems.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Determination of Crystal Structures
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Ultrasonography
During an ultrasonography procedure, a handheld device called...