Validation of the Oncentra Brachy Advanced Collapsed cone Engine for a commercial (192)Ir source using heterogeneous
Yunzhi Ma1, Fréderic Lacroix2, Marie-Claude Lavallée2
1Département de physique, de génie physique et d'optique et Centre de recherche sur le cancer, Université Laval, Québec, Canada; Département de radio-oncologie et CRCHU de Québec, CHU de Québec, Québec, Canada.
Purpose:
To validate the Advanced Collapsed cone Engine (ACE) dose calculation engine of Oncentra Brachy (OcB) treatment planning system using an (192)Ir source.
Methods And Materials:
Two levels of validation were performed, conformant to the model-based dose calculation algorithm commissioning guidelines of American Association of Physicists in Medicine TG-186 report. Level 1 uses all-water phantoms, and the validation is against TG-43 methodology. Level 2 uses real-patient cases, and the validation is against Monte Carlo (MC) simulations. For each case, the ACE and TG-43 calculations were performed in the OcB treatment planning system. ALGEBRA MC system was used to perform MC simulations.
Results:
In Level 1, the ray effect depends on both accuracy mode and the number of dwell positions. The volume fraction with dose error ≥2% quickly reduces from 23% (13%) for a single dwell to 3% (2%) for eight dwell positions in the standard (high) accuracy mode. In Level 2, the 10% and higher isodose lines were observed overlapping between ACE (both standard and high-resolution modes) and MC. Major clinical indices (V100, V150, V200, D90, D50, and D2cc) were investigated and validated by MC. For example, among the Level 2 cases, the maximum deviation in V100 of ACE from MC is 2.75% but up to ~10% for TG-43. Similarly, the maximum deviation in D90 is 0.14 Gy between ACE and MC but up to 0.24 Gy for TG-43.
Conclusion:
ACE demonstrated good agreement with MC in most clinically relevant regions in the cases tested. Departure from MC is significant for specific situations but limited to low-dose (<10% isodose) regions.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...