Safety Profile of a Miniaturized Insertable Cardiac Monitor: Results from Two Prospective Trials
Related Experiment Video
Updated: Apr 2, 2026

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Suneet Mittal1, Prashanthan Sanders2, Evgeny Pokushalov3
1Electrophysiology Laboratory, The Valley Hospital, Valley Health System, Ridgewood, New Jersey.
Background:
Insertable cardiac monitors (ICMs) are used to continuously monitor the patient's electrocardiogram. In response to patient activation or based on automated device algorithms, arrhythmia episodes are stored and automatically transmitted daily to the clinician. Thus, ICMs can be used to diagnose arrhythmias in at-risk patients and in those with symptoms potentially attributable to arrhythmias. The ICM described in this report has undergone a dramatic change in size and method of insertion.
Methods:
To evaluate the safety profile of the ICM procedure, we analyzed procedure-related adverse events (AEs) from two separate trials: A controlled, nonrandomized multicenter study (Reveal LINQ(TM) Usability study) and a multicenter registry (Reveal LINQ(TM) Registry) evaluating real-world experience. For the Registry we reported all procedure-related AEs upon occurrence, whereas for the Usability study, we reported events occurring during the first month of follow-up.
Results:
The Usability study enrolled 151 patients (age 56.6 ± 12.1 years; male 67%) at 16 centers; during follow-up, an infection was observed in 1.3% patients and a procedure-related serious AE (SAE) in 0.7% patients. The Registry enrolled 122 patients (age 61.0 ± 17.8 years; male 47%) at seven centers; during follow-up, an infection was observed in 1.6% patients and a procedure-related SAE in 1.6% patients.
Conclusions:
The cumulative experience from a controlled clinical trial and a "real-world" registry demonstrate that the new ICM can be inserted with very low incidence of AEs.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Imaging Studies for Cardiovascular System IV: CMRI