Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
4.6K
Fragmented red cells reference range (Sysmex XN(®) automated blood cell counter)
Jean-François Lesesve1, Amandine Daigney1, Sylvain Henry1
1Service d'hématologie biologique, CHU Nancy, France.
Annales De Biologie Clinique
|October 23, 2015
Summary
Fragmented red cells (FRCs) are a new automated blood count parameter. Establishing a device-specific reference range for FRCs on the Sysmex XN-9000 is crucial for accurate schistocyte assessment.
Area of Science:
- Hematology
- Clinical Pathology
- Laboratory Medicine
Background:
- Fragmented red cells (FRCs) are a novel parameter from automated blood cell counters.
- FRCs are considered indicators of schistocytes, which are typically assessed via manual microscopy.
- Automated FRC counts are influenced by the detection methodology, necessitating device-specific reference ranges.
Purpose of the Study:
- To establish a reference range for Fragmented Red Cells (FRCs) using the Sysmex XN-9000 automated hematology analyzer.
- To investigate the correlation between FRCs and red blood cells with low hemoglobin content (Hypo-He).
Main Methods:
- Analysis of 2389 control blood samples using the Sysmex XN-9000 analyzer.
- Statistical calculation of mean, standard deviation, and median for FRC percentages.
- Correlation analysis between FRC counts and Hypo-He percentages.
Main Results:
- The established reference range for FRCs on the Sysmex XN-9000 was a mean of 0.32% ± 0.81, with a median of 0.02%.
- A significant correlation was observed between increased FRC counts and elevated percentages of Hypo-He red blood cells.
- Hypo-He was identified as a factor contributing to spurious overestimation of FRCs.
Conclusions:
- The derived reference range for FRCs on the Sysmex XN-9000 aids laboratory staff in identifying blood smears requiring manual review.
- This data can assist Sysmex in optimizing automated rules for blood smear analysis and automated slide making.
- Accurate FRC quantification is essential for reliable schistocyte detection and interpretation in automated hematology.