[Readspace solid-phase microextraction-gas chromatography for determination of 2,5-hexanedione in urine]
Objective:
To establish a method for determination of 2,5-hexanedione in urine by headspace solid-phase microextraction-gas chromatography.
Methods:
After extraction by solid-phase microextraction head, 2,5-hexanedione in urine was determined by gas chromatography and was quantified by external standard method.
Results:
The concentration of 2,5-hexanedione in urine showed a linear relationship within the range of 0.1-20.0 µg/ml. The regression equation was y=261.36x-1.903 3, r=0.999 2. The minimum detectable concentration was 0.01 µg/ml. The recovery rate was 92.6%-97.1%, with a relative standard deviation (RSD) of 3.3%-5.8%. The intra-day and inter-day RSDs were 3.8%-6.2% and 4.7%-6.3% respectively.
Conclusion:
This determination method has no requirement for organic solvents, features simple and rapid operation, possesses higher detection sensitivity, and applies well to the determination of 2,5-hexanedione in urine.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
11:06GC-based Detection of Aldononitrile Acetate Derivatized Glucosamine and Muramic Acid for Microbial Residue Determination in Soil
Published on: May 19, 2012
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...