A liposome-based size calibration method for measuring microvesicles by flow cytometry
1Department of Chemistry and Nano-Science Center, University of Copenhagen, Copenhagen, Denmark.
Journal of Thrombosis and Haemostasis : JTH
|October 29, 2015
Summary
Accurate microvesicle sizing by flow cytometry requires a gold standard. Fluorescent liposomes serve as superior size calibrators compared to polystyrene beads, enhancing microvesicle size assessment.
Area of Science:
- Extracellular vesicle research
- Flow cytometry applications
- Nanoparticle characterization
Background:
- Accurate sizing of microvesicles is crucial for understanding their biological roles.
- Existing methods for microvesicle size determination by flow cytometry lack a standardized gold standard.
- The need for reliable calibrators in flow cytometry-based vesicle analysis is increasingly recognized.
Purpose of the Study:
- To establish a more accurate method for determining microvesicle sizes using flow cytometry.
- To evaluate the suitability of fluorescently labeled liposomes as size calibrators.
- To compare the performance of liposome-based calibrators against traditional polystyrene beads.
Main Methods:
- Preparation of fluorescently labeled liposomes with defined sizes (200-1000 nm) via extrusion.
- Measurement of side scattering (SSC) signals from liposomes and microvesicles using flow cytometry.
- Correlation analysis between liposome size, fluorescence intensity, and SSC signals.
Main Results:
- Fluorescent liposomes demonstrated a reliable correlation between SSC and vesicle size.
- Polystyrene beads showed significant inaccuracies, with 110 nm beads correlating to ~400 nm vesicles.
- The SSC-based size assessment method exhibits a resolution limit of approximately 60-200 nm.
Conclusions:
- Fluorescent liposomes offer a more accurate size calibration method for microvesicles in flow cytometry.
- Liposome-based calibration improves the reliability of microvesicle size estimation compared to polystyrene beads.
- The study highlights the limitations and potential improvements for flow cytometry-based vesicle sizing.